BQS Contact Information Fax - (303) 236 1880 |
Long Term-Method Detection Levels (LT-MDL) |
2009-2010 (spiking starts May 1, 2008 |
||
2010-2011 |
||
2007-2008 (in use) |
2011-2012 |
|
2008-2009 (review in process) |
2012-2013 |
|
|
|
Formula
|

LT-MDL = s x t
Where:
s = standard deviation of, ideally, 24 samples (no less than 7)
t = student’s t value for n-1 observations
The LT-MDL is set to limit the false positives to less than or equal to
one percent. The LT-MDL is always set based on hind-sight. In other words, the
previous year ’s data are used to determine the LT-MDL to use this year. This
can be frustrating when past events limit current flexibility. However, in an
effort to minimize the number of times an LT-MDL might change in its lifetime,
the once per year assessment was determined to be optimal for data users.
Setting the LT-MDL too high will over protect
against false positives so that more real detections are censored. Setting the
LT-MDL too low will result in more environmental sample results released that
are due to laboratory background and not from the environmental sample itself.
The assumption is that the standard deviation of blank values, times the
student ’s t value will result in the
concentration equal to about the 99th percentile of the blanks. This assumption relies on
a Normal Distribution.
|
|
The purpose of using a spiked sample to calculate the LT-MDL is to overcome the limitation that some methods do not produce signal for blank results. Without knowing the blank population, it is not possible to know where false positives might be more likely to occur. The theory is that a very low concentration spike will have essentially the same distribution as a blank, so very low concentration spike samples can be used to determine the LT-MDL of analyses in the absence of blanks. For information rich analyses (such as mass spectometer methods) the LT-MDL is most often associated with the instrument sensitivity, and not the presence of interferences. More work is being done to investigate better ways to identify the detection level for information rich methods.
You must spike at a concentration where the distribution of results is very similar to the expected blank concentration. The spike concentration should be high enough to be reliably detectable (otherwise you won ’t have any data) and low enough to be able to assume a similar distribution of results as the blank. This is accepted to be about 2 – 5 times the expected LT-MDL.
n alternate process used to assure that the LT-MDL appropriately accounts for blank detections is to use the “Blank-corrected LT-MDL” for methods that produce blank signals. Here the LT-MDL is based on the original goal of setting the censoring level at less-than 1% false positives. To do this, in a set of 100 blank results, the LT-MDL would be set at the 99th highest ranked value (the second highest). We typically use data sets of around 40 blanks, so setting the LT-MDL equal to the second highest ranked blank value is not as conservative an approach, but an easy process that is easy to visualize and implement.
Setting the LTMDL at
the 1% F+ level
| Outlier Corrections |
To remove true outliers, the first line of assurance is to request reanalysis of LT-MDL or Blind Blank samples that fall outside of plus or minus two times the LT-MDL in use. The laboratory will only replace a sample result if the new result is significantly different than the first, and verified.
Then, an outlier
test is applied - the Grubbs Test. This is used to
eliminate only the most aberrant single data points that individually skew a
data set. The Grubbs test will check for
instances where a single (or very few) data point(s) are not known to be in
error (a rerun produced the same result), and are obviously completely outside
of the norm. The reason you would want to apply this filter to the data is because
this single aberrant point could increase the LT-MDL so much that 99% of the
lab’s customers would be negatively affected by overly-censored data, to save
1% of the customers from a false positive. Granted, this decision is a matter
of choice.
|
|
|
Updating the LT-MDLs |
LT-MDL values are stored with only one significant digit. Reporting levels are stored with two significant digits.
|
|

Alternate data reporting schemes beyond the LT-MDL and LRL approach are the method reporting level (MRL) and the interim reporting level (IRL).
The IRL is used when a method does not have at least one year's worth of supporting data to determine an LT-MDL and an LRL. The IRL is based on an estimated detection limit, and is set at least two times that estimation. Data between the IRL and the detection limit may be reported for information rich methods such as mass spectrometry, but detections less than the detection limit are censored in the LIMs and are not available to customers.