- Aluminum, FIL, ICPMS -
Blind Blank Results
Test ID: 01106PLM43
Open Data Set

The SGPlot Procedure

False Positive Rate for Aluminum, FIL, ICPMS

Analyte Test ID Lab Code Water Year # of
Samples
# of
False Pos.
*False
Positive Rate
Aluminum, FIL, ICPMS 01106PLM43 1784 2015 52 4 8%
      2016 51 5 10%
      2017 3 0 0%
*Highlighted if false positive rate > 5 %

Updated 02/17/2017



- Antimony, FIL, ICPMS -
Blind Blank Results
Test ID: 01095PLM43
Open Data Set

The SGPlot Procedure

False Positive Rate for Antimony, FIL, ICPMS

Analyte Test ID Lab Code Water Year # of
Samples
# of
False Pos.
*False
Positive Rate
Antimony, FIL, ICPMS 01095PLM43 1785 2015 52 0 0%
      2016 51 3 6%
      2017 3 0 0%
*Highlighted if false positive rate > 5 %

Updated 02/17/2017



- Arsenic, FIL, cICPMS -
Blind Blank Results
Test ID: 01000PLM10
Open Data Set

The SGPlot Procedure

False Positive Rate for Arsenic, FIL, cICPMS

Analyte Test ID Lab Code Water Year # of
Samples
# of
False Pos.
*False
Positive Rate
Arsenic, FIL, cICPMS 01000PLM10 3122 2015 52 0 0%
      2016 51 1 2%
      2017 3 0 0%
*Highlighted if false positive rate > 5 %

Updated 02/17/2017



- Barium, FIL, ICPMS -
Blind Blank Results
Test ID: 01005PLM43
Open Data Set

The SGPlot Procedure

False Positive Rate for Barium, FIL, ICPMS

Analyte Test ID Lab Code Water Year # of
Samples
# of
False Pos.
*False
Positive Rate
Barium, FIL, ICPMS 01005PLM43 1786 2015 52 0 0%
      2016 51 4 8%
      2017 3 0 0%
*Highlighted if false positive rate > 5 %

Updated 02/17/2017



- Beryllium, FIL, ICPMS -
Blind Blank Results
Test ID: 01010PLM43
Open Data Set

The SGPlot Procedure

False Positive Rate for Beryllium, FIL, ICPMS

Analyte Test ID Lab Code Water Year # of
Samples
# of
False Pos.
*False
Positive Rate
Beryllium, FIL, ICPMS 01010PLM43 1787 2015 52 0 0%
      2016 51 1 2%
      2017 3 0 0%
*Highlighted if false positive rate > 5 %

Updated 02/17/2017



- Boron, FIL, ICPMS -
Blind Blank Results
Test ID: 01020PLM40
Open Data Set

The SGPlot Procedure

False Positive Rate for Boron, FIL, ICPMS

Analyte Test ID Lab Code Water Year # of
Samples
# of
False Pos.
*False
Positive Rate
Boron, FIL, ICPMS 01020PLM40 2504 2015 52 0 0%
      2016 51 1 2%
      2017 3 0 0%
*Highlighted if false positive rate > 5 %

Updated 02/17/2017



- Cadmium, FIL, ICPMS -
Blind Blank Results
Test ID: 01025PLM43
Open Data Set

The SGPlot Procedure

False Positive Rate for Cadmium, FIL, ICPMS

Analyte Test ID Lab Code Water Year # of
Samples
# of
False Pos.
*False
Positive Rate
Cadmium, FIL, ICPMS 01025PLM43 1788 2015 52 0 0%
      2016 51 1 2%
      2017 3 0 0%
*Highlighted if false positive rate > 5 %

Updated 02/17/2017



- Chromium, FIL, cICPMS -
Blind Blank Results
Test ID: 01030PLM10
Open Data Set

The SGPlot Procedure

False Positive Rate for Chromium, FIL, cICPMS

Analyte Test ID Lab Code Water Year # of
Samples
# of
False Pos.
*False
Positive Rate
Chromium, FIL, cICPMS 01030PLM10 3126 2015 52 0 0%
      2016 51 0 0%
      2017 3 0 0%
*Highlighted if false positive rate > 5 %

Updated 02/17/2017



- Cobalt, FIL, cICPMS -
Blind Blank Results
Test ID: 01035PLM10
Open Data Set

The SGPlot Procedure

False Positive Rate for Cobalt, FIL, cICPMS

Analyte Test ID Lab Code Water Year # of
Samples
# of
False Pos.
*False
Positive Rate
Cobalt, FIL, cICPMS 01035PLM10 3124 2015 52 0 0%
      2016 51 1 2%
      2017 3 0 0%
*Highlighted if false positive rate > 5 %

Updated 02/17/2017



- Copper, FIL, cICPMS -
Blind Blank Results
Test ID: 01040PLM10
Open Data Set

The SGPlot Procedure

False Positive Rate for Copper, FIL, cICPMS

Analyte Test ID Lab Code Water Year # of
Samples
# of
False Pos.
*False
Positive Rate
Copper, FIL, cICPMS 01040PLM10 3128 2015 52 1 2%
      2016 51 2 4%
      2017 3 0 0%
*Highlighted if false positive rate > 5 %

Updated 02/17/2017



- Lead, FIL, ICPMS -
Blind Blank Results
Test ID: 01049PLM43
Open Data Set

The SGPlot Procedure

False Positive Rate for Lead, FIL, ICPMS

Analyte Test ID Lab Code Water Year # of
Samples
# of
False Pos.
*False
Positive Rate
Lead, FIL, ICPMS 01049PLM43 1792 2015 52 1 2%
      2016 51 3 6%
      2017 3 0 0%
*Highlighted if false positive rate > 5 %

Updated 02/17/2017



- Lithium, FIL, ICPMS -
Blind Blank Results
Test ID: 01130PLM40
Open Data Set

The SGPlot Procedure

False Positive Rate for Lithium, FIL, ICPMS

Analyte Test ID Lab Code Water Year # of
Samples
# of
False Pos.
*False
Positive Rate
Lithium, FIL, ICPMS 01130PLM40 2505 2015 52 0 0%
      2016 73 1 1%
      2017 5 0 0%
*Highlighted if false positive rate > 5 %

Updated 02/17/2017



- Manganese, FIL, ICPMS -
Blind Blank Results
Test ID: 01056PLM43
Open Data Set

The SGPlot Procedure

False Positive Rate for Manganese, FIL, ICPMS

Analyte Test ID Lab Code Water Year # of
Samples
# of
False Pos.
*False
Positive Rate
Manganese, FIL, ICPMS 01056PLM43 1793 2015 52 0 0%
      2016 51 1 2%
      2017 3 0 0%
*Highlighted if false positive rate > 5 %

Updated 02/17/2017



- Molybdenum, FIL, ICPMS -
Blind Blank Results
Test ID: 01060PLM43
Open Data Set

The SGPlot Procedure

False Positive Rate for Molybdenum, FIL, ICPMS

Analyte Test ID Lab Code Water Year # of
Samples
# of
False Pos.
*False
Positive Rate
Molybdenum, FIL, ICPMS 01060PLM43 1794 2015 52 0 0%
      2016 58 2 3%
      2017 5 0 0%
*Highlighted if false positive rate > 5 %

Updated 02/17/2017



- Nickel, FIL, cICPMS -
Blind Blank Results
Test ID: 01065PLM10
Open Data Set

The SGPlot Procedure

False Positive Rate for Nickel, FIL, cICPMS

Analyte Test ID Lab Code Water Year # of
Samples
# of
False Pos.
*False
Positive Rate
Nickel, FIL, cICPMS 01065PLM10 3130 2015 52 0 0%
      2016 58 5 9%
      2017 5 1 20%
*Highlighted if false positive rate > 5 %

Updated 02/17/2017



- Selenium, FIL, cICPMS -
Blind Blank Results
Test ID: 01145PLM10
Open Data Set

The SGPlot Procedure

False Positive Rate for Selenium, FIL, cICPMS

Analyte Test ID Lab Code Water Year # of
Samples
# of
False Pos.
*False
Positive Rate
Selenium, FIL, cICPMS 01145PLM10 3132 2015 52 0 0%
      2016 51 1 2%
      2017 3 0 0%
*Highlighted if false positive rate > 5 %

Updated 02/17/2017



- Silver, FIL, ICPMS -
Blind Blank Results
Test ID: 01075PLM43
Open Data Set

The SGPlot Procedure

False Positive Rate for Silver, FIL, ICPMS

Analyte Test ID Lab Code Water Year # of
Samples
# of
False Pos.
*False
Positive Rate
Silver, FIL, ICPMS 01075PLM43 1796 2015 52 1 2%
      2016 51 1 2%
      2017 3 0 0%
*Highlighted if false positive rate > 5 %

Updated 02/17/2017



- Strontium, FIL, ICPMS -
Blind Blank Results
Test ID: 01080PLM40
Open Data Set

The SGPlot Procedure

False Positive Rate for Strontium, FIL, ICPMS

Analyte Test ID Lab Code Water Year # of
Samples
# of
False Pos.
*False
Positive Rate
Strontium, FIL, ICPMS 01080PLM40 2507 2015 52 0 0%
      2016 51 2 4%
      2017 3 0 0%
*Highlighted if false positive rate > 5 %

Updated 02/17/2017



- Thallium, FIL, ICPMS -
Blind Blank Results
Test ID: 01057PLM40
Open Data Set

The SGPlot Procedure

False Positive Rate for Thallium, FIL, ICPMS

Analyte Test ID Lab Code Water Year # of
Samples
# of
False Pos.
*False
Positive Rate
Thallium, FIL, ICPMS 01057PLM40 2508 2015 52 0 0%
      2016 51 1 2%
      2017 3 0 0%
*Highlighted if false positive rate > 5 %

Updated 02/17/2017



- Tungsten, FIL, ICPMS -
Blind Blank Results
Test ID: 01155PLM10
Open Data Set

The SGPlot Procedure

False Positive Rate for Tungsten, FIL, ICPMS

Analyte Test ID Lab Code Water Year # of
Samples
# of
False Pos.
*False
Positive Rate
Tungsten, FIL, ICPMS 01155PLM10 3136 2015 52 0 0%
      2016 51 0 0%
      2017 3 0 0%
*Highlighted if false positive rate > 5 %

Updated 02/17/2017



- Uranium, FIL, ICPMS -
Blind Blank Results
Test ID: 22703PLM43
Open Data Set

The SGPlot Procedure

False Positive Rate for Uranium, FIL, ICPMS

Analyte Test ID Lab Code Water Year # of
Samples
# of
False Pos.
*False
Positive Rate
Uranium, FIL, ICPMS 22703PLM43 1797 2015 52 0 0%
      2016 51 2 4%
      2017 3 0 0%
*Highlighted if false positive rate > 5 %

Updated 02/17/2017



- Vanadium, FIL, cICPMS -
Blind Blank Results
Test ID: 01085PLM10
Open Data Set

The SGPlot Procedure

False Positive Rate for Vanadium, FIL, cICPMS

Analyte Test ID Lab Code Water Year # of
Samples
# of
False Pos.
*False
Positive Rate
Vanadium, FIL, cICPMS 01085PLM10 3134 2015 52 3 6%
      2016 51 1 2%
      2017 3 0 0%
*Highlighted if false positive rate > 5 %

Updated 02/17/2017



- Zinc, FIL, cICPMS -
Blind Blank Results
Test ID: 01090PLM10
Open Data Set

The SGPlot Procedure

False Positive Rate for Zinc, FIL, cICPMS

Analyte Test ID Lab Code Water Year # of
Samples
# of
False Pos.
*False
Positive Rate
Zinc, FIL, cICPMS 01090PLM10 3138 2015 52 1 2%
      2016 51 2 4%
      2017 3 0 0%
*Highlighted if false positive rate > 5 %

Updated 02/17/2017